L-LAS-CAM Series
Line camerasGeneral technical data ....
GENERAL TECHNICAL DATA:
Software | Windows® PC software: L-LAS-CAM-Scope |
Voltage supply | +24VDC (± 10%), reversed polarity protected, overload protected |
Current consumption | <200mA |
Max. switching current | Digital outputs OUT0, OUT1: 100mA, short-circuit proof |
Line detector | depends on sensor type |
Reference distance | depends on measuring objective used |
Working range (measuring distance) |
depends on measuring objective used |
Resolution | 0.4% of measuring range |
Light source (optional) |
Various external lighting units are available (LU-LED-...-VIS, LU-LED-...-VIS-DIF) |
Analog output | Voltage 0...+10V or current 4...20mA (can be adjusted under Windows® on PC) |
Digital outputs | OUT0, OUT1, OUT2: pnp bright-switching/npn dark-switching or pnp dark-switching/npn bright-switching output polarity can be adjusted under Windows® on PC |
Digital inputs | IN0: External trigger IN1: Teach/Reset (double function) Input voltage +Ub/0V, with protective circuit |
Sensitivity | adjustable under Windows® on PC |
Teach/Reset button | for norm value teaching or for reset of maximum values |
Intensity adjustment | adjustable under Windows® on PC |
LED display | LED red (+): Measuring value > upper tolerance threshold LED green: Measuring value within tolerance window LED red (-): Measuring value < lower tolerance threshold LED yellow: LED for sensor adjustment (multifunctional) LED yellow: POWER-LED |
Housing dimensions | depends on sensor type |
Housing material | Aluminum, anodized in black |
Type of connector | - to PLC: 8-pole fem. connector Binder 712 - to PC: 4-pole fem. connector Binder 707 - to external lighting unit: 4-pole fem. connector Binder 768 |
Connecting cables | - to PLC: cab-las8/SPS - to PC: cab-las4/PC, cab-4/USB or cab-4/ETH - to external lighting unit LU-LED-...: cab-slu4/768-fem |
Operating temp. range | -10°C ... +60°C |
Storage temp. range | -20°C ... +85°C |
Enclosure rating | IP54 |
EMC test acc. to | DIN EN 60947-5-2 |
Available measuring objectives |
MO-C-1,4/12,5mm MO-C-1,4/16mm MO-C-1,4/25mm MO-C-1,4/35mm MO-C-1,8/50mm MO-C-1,8/75mm |
External lighting units (optional) |
LU-LED-110-VIS (length of illuminated area: 100 mm) LU-LED-430-VIS (length of illuminated area: 430 mm) LU-LED-870-VIS (length of illuminated area: 870 mm) LU-LED-1470-VIS (length of illuminated area: 1470 mm) LU-LED-870-VIS-DIF (length of illuminated area: 870 mm) LU-LED-1470-VIS-DIF (length of illuminated area: 1470 mm) |
Filter inserts (optional) |
FILT-C-9-3-IF670 (suitable for measuring objects MO-C-...) FILT-C-9-3-GG475 (suitable for measuring objects MO-C-...) FILT-C-9-3-KG5 (suitable for measuring objects MO-C-...) FILT-C-9-3-RG715 (suitable for measuring objects MO-C-...) FILT-C-17-3-IF670 (suitable for measuring objectse MO-C-...) FILT-C-17-3-GG475 (suitable for measuring objects MO-C-...) FILT-C-17-3-KG5 (suitable for measuring objects MO-C-...) FILT-C-17-3-RG715 (suitable for measuring objects MO-C-...) FILT-C-26-3-GG475 (suitable for measuring objectse MO-C-...) Tool for mounting of optical filters FILT-C: MOUNT-FILT |
Intermediate tubes/ adapter rings (optional) |
TUB-10 (suitable for all measuring objects MO-...) TUB-20 (suitable for all measuring objects MO-...) TUB-30 (suitable for all measuring objects MO-...) TUB-50 (suitable for all measuring objects MO-...) TUB-100 (suitable for all measuring objects MO-...) |